Keyence pressure sensors, in six range variants.

Pressure and temperature — see Pressure. Models GP-M001T, GP-M010T, GP-M025T, GP-M100T, GP-M250T, and GP-M400T; the range depends on the model.
| Model | Rated pressure | Display range | Allowable pressure | Burst pressure | Display resolution |
|---|---|---|---|---|---|
| GP-M001T (compound) | −100 to +100 kPa (−14.50 to +14.50 psi) | −120.0 to +120.0 kPa (−17.40 to +17.40 psi) | 400 kPa (58 psi) | 1.5 MPa (218 psi) | 0.1 kPa (0.015 psi) |
| GP-M010T | −0.1 to +1 MPa (−14.5 to +145.0 psi) | −0.210 to +1.110 MPa (−30.5 to +161.0 psi) | 4 MPa (580 psi) | 15 MPa (2176 psi) | 0.001 MPa (0.15 psi) |
| GP-M025T | −0.1 to +2.5 MPa (−14.5 to +362.6 psi) | −0.360 to +2.760 MPa (−52.2 to +400.3 psi) | 10 MPa (1450 psi) | 35 MPa (5076 psi) | 0.001 MPa (0.15 psi) |
| GP-M100T | 0 to +10 MPa (0 to +1450 psi) | −1.00 to +11.00 MPa (−145.0 to +1595.4 psi) | 30 MPa (4351 psi) | 100 MPa (14504 psi) | 0.01 MPa (1.5 psi) |
| GP-M250T | 0 to +25 MPa (0 to +3626 psi) | −2.50 to +27.50 MPa (−362.6 to +3988.5 psi) | 50 MPa (7252 psi) | 100 MPa (14504 psi) | 0.01 MPa (1.5 psi) |
| GP-M400T | 0 to +40 MPa (0 to +5802 psi) | −4.00 to +44.00 MPa (−580.1 to +6381.7 psi) | 50 MPa (7252 psi) | 100 MPa (14504 psi) | 0.01 MPa (1.5 psi) |
All six measure gauge pressure. Common to all variants:
| Accuracy | ±1.0 % of F.S. or less |
| Repeatability | ±0.3 % of F.S. or less |
| Temperature characteristics | ±0.6 % of F.S. / 10 °C |
| Response time | 3 to 5000 ms, settable |
| Built-in temperature sensor | Display range −32 to +112 °C (−25.6 to +233.6 °F); accuracy ±3.5 °C (±6.3 °F) at 25 °C ambient (as listed for the GP-M001T) |
| Media | Gas or liquid that will not corrode the wetted parts |
| Supply | 10–30 V DC, ripple (P-P) 10 % max., Class 2 or LPS |
| Current consumption | 60 mA or less, excluding output (at 24 V: 30 mA or less; at 12 V: 50 mA or less) |
| Output | IO-Link on pin 4 (output 1 when no IO-Link master is connected) |
| Switch outputs | 2; NPN/PNP open collector, selectable; N.O./N.C. selectable. Output 2 (pin 2) can instead be set as the analog output |
| Switching output rating | 30 V or less, max. 100 mA, residual voltage 1 V or less |
| Analog output | 4–20 mA, max. load resistance 260 Ω (on pin 2, when selected) |
| IO-Link | v1.1, COM2 (38.4 kbps), min. cycle time 4.3 ms |
| Connector | M12, 4-pin (pin 1 brown 10–30 V DC, pin 2 white output 2 / analog, pin 3 blue 0 V, pin 4 black output 1 / IO-Link) |
| Ingress rating | IP67 |
| Ambient temperature | −20 to +80 °C (−4 to +176 °F), no freezing or condensation |
| Medium temperature | −20 to +100 °C (−4 to +212 °F), no freezing or condensation |
| Process connection | G3/4; changes to R1/8 male, R1/4 male, R3/8 male, G1/4 female, NPT1/8 male or NPT1/4 male with Keyence adapters |
| Wetted materials | Pressure port SUS XM7; diaphragm Al₂O₃ (ceramic); O-ring FKM |
| Housing materials | Metal: SUS304, SUS303; plastic: PPSU |
| Weight | Approx. 150 g |
Wiring and configuration live with each controller: