A Pulsafeeder toroidal (electrodeless) conductivity probe for use with Pulsafeeder controllers.

Some ratings are not filled in yet. Check the manufacturer’s current datasheet before wiring.
Conductivity — see Conductivity.
| Model | From datasheet — to be filled in |
| Output | Connects to a MicroVision EX probe input; the conductivity probe is prewired to Probe 1 (make-up conductivity to Probe 4) |
| Controller reading range | 0–9,999 µS/cm in 1 µS/cm steps |
| Temperature compensation | Built in; compensation range 0–50 °C (32–122 °F) |
| Process connection | From datasheet — to be filled in |
| Pressure and temperature limits | 125 psi (8.6 bar), 50 °C (122 °F) maximum |
| Materials | Polypropylene |
| Maximum cable length | 30.5 m (100 ft) |
The compensation, pressure, temperature, material and cable figures are Pulsafeeder’s specification for the toroidal probe supplied with the MicroVision EX. Pulsafeeder supplies it already installed in a tee. It works by driving an AC current through one coil, which induces a current in the water and, in turn, in a second pick-up coil; the pick-up current is proportional to conductivity.
A toroidal probe has no metal electrodes to foul, so it needs no electrode cleaning and no routine recalibration.
Wiring and configuration live with each controller: