The NextStep dosing pump reports an over-temperature condition.
| At a glance | |
|---|---|
| Code | 335 |
| Shown as | Pump Over-Temperature |
| In the alarm list | The device’s name, then Over-Temperature — for example Acid Pump Over-Temperature (upcoming firmware) |
| Severity | Severe |
| Category | 3xx — Process / sensor signals |
| Clears | Automatically, when the pump clears the condition |
The pump raised this condition itself; the controller reads it from the pump’s alarm register and shows it as an alarm. It is keyed to the pump, so the alarm names which pump raised it.
Records the alarm. The pump acts on its own alarms; the controller does not change the pump’s command because of it.
Inspect the pump and its surroundings — ventilation, direct sun, a stalled motor.