Calibrating a probe from the controller, so its readings match a reference.
Open the probe from Devices, press Enter for its options, then choose Calibrate. The steps depend on the probe.
New to calibration? Read this first. If you have done it before, skip to your probe.
A probe’s raw signal drifts as it ages and fouls. Calibration compares the probe with something you trust — a standard solution, or a lab or handheld reading — and corrects the probe so the two agree. It does not clean the probe or fix a failing one: it hides a small error, and a large one means the probe needs attention.
A probe can be wrong in two ways:
A one-point calibration corrects one of the two. On a conductivity probe it corrects the slope (the zero is set by Air Cal); on a pH probe it corrects the offset and keeps the slope. A two-point or three-point calibration corrects both, and is what the Pyxis pH probes run (pH 7, 4 and 10).
Process calibration leaves the probe in the line and enters a reading taken by a trusted meter or a lab. Control keeps running. It is quick and good for a routine check, but only as good as the meter or the sample.
Buffer calibration takes the probe out and puts it in a solution of known value. It is more accurate, and it is the only way to check a probe you suspect, but control is paused while the probe is out.
A good routine is a process check every visit and a buffer calibration every month or two, or whenever the process check is more than a little off.
The controller decides for you. At each step it waits until the reading stays within a small band for a set time — for example 0.05 pH for 30 seconds — and shows its progress while it waits. A probe that will not settle is telling you something: air bubbles, a dirty sensor, a buffer at a different temperature from the probe, or electrical noise. Fix the cause rather than pressing on.
Clean first, always, if the probe looks dirty or has been in the line a month or more. Calibrating a fouled probe matches it to the standard today, and it is wrong again as soon as the fouling changes. The controller refuses a correction that is too large (Gain out of range, Offset out of range): that is almost always a probe that needs cleaning, a wrong or old standard, or a probe at the end of its life.
A process calibration does not. A buffer calibration, and every Pyxis calibration, holds the control modes that use the probe from the moment you take it out until you finish — see each procedure below. Plan for the time the system runs without that mode.
| Probe | Calibration | Where the correction is kept |
|---|---|---|
| Sensorex Smart Sensor, conductivity | Process, or buffer with air zero and one standard | In the controller |
| Sensorex Smart Sensor, pH (upcoming) | Process, or buffer at one standard | In the controller |
| Pyxis ST-500 (upcoming) | Slope in a PTSA standard, or zero in DI water, after a cleanliness check | In the probe |
| Pyxis ST-710 (upcoming) | pH 7, 4 and 10 | In the probe |
| Pyxis ST-712 (upcoming) | pH 7, 4 and 10; or ORP at one standard | In the probe |
| Pyxis CR-300 (upcoming) | Check cap | In the probe |
In the controller means the probe’s own reading is untouched, and the controller corrects it. In the probe means the controller runs the probe’s own calibration procedure and stores nothing of its own. A probe never carries both corrections: calibrating an ST-712 at pH 7 clears any correction the controller held for it from older firmware.
New in an upcoming firmware update. Earlier firmware calibrates the Sensorex conductivity probe only. Other probes are calibrated per their manufacturer’s instructions.
Only one calibration can run at a time. If another is open, the controller says Calibration unavailable. Is one running?
| Method | The probe | Use it when | Control |
|---|---|---|---|
| Process | Stays installed, in the water | Checking against a handheld meter or a lab result | Keeps running |
| Buffer Cal | Comes out, into a known standard | A full calibration, or the probe is badly off | Paused for the duration |


The probe never leaves the line, so control keeps running throughout.
Buffer calibration pauses control. Any control mode using this probe stops while the probe is out of the water, and stays stopped until you resume it at the end. Plan for the time the system will run without it.


A one-point pH calibration shifts the reading to match the standard and keeps the probe’s slope. A two-point pH calibration is not offered yet.
Sensorex pH calibration is not yet tested on hardware. Check the reading against a buffer after calibrating.
Exit leaves the calibration. No correction is saved until you press Apply.
At each step the controller waits for the reading to settle, then advances by itself. The screen shows the live reading while it waits. If a probe will not settle, you can move on anyway.
| Probe | Steady within | For |
|---|---|---|
| Conductivity | 2 µS/cm | 30 seconds |
| pH | 0.05 pH | 30 seconds |
The Review step shows Failed, the reason, and what to try — for example Clean probe, retry. Choose Retry to go again.
| Message | Means | Do |
|---|---|---|
| Gain out of range | The correction needed is too large. Conductivity: more than twice, or less than half, the raw reading. | Check the standard’s value, clean the probe, and retry. A probe that still fails needs service. |
| Offset out of range | pH: the correction needed is more than 2.4 pH either way | Check the buffer is fresh and is the value you entered. A probe this far off needs cleaning or replacing. |
| Reading unusable | No valid reading to calibrate against | Check the probe is connected and reading. |
| Not saved — Lost on restart. Retry to save. | The calibration was applied but could not be stored | Retry, or it will be lost at the next restart. |
Reset Values returns the probe to its factory calibration, now and after a restart.
The Pyxis probes calibrate themselves. The controller walks you through each probe’s own procedure, sends the probe its commands, and shows the probe’s verdict in words.
Control is held from the Prepare step until you leave. Once you press Ready to put the probe in the standard, every control mode using the probe is held. The hold ends when you leave the calibration — after Done, or Back before anything was sent.
The buttons are the same at every step: Up and Down change the value, Enter moves on, Back goes back a step. While the probe is working (Checking…, Calibrating…) the buttons wait for it.
The ST-500 calibrates its PTSA reading. It checks its own optics first, so a dirty probe is caught before it is calibrated.

Keep direct light off the probe while it settles. Pyxis’s own guidance is in Pyxis ST-500 — Maintenance.
| Result | Do |
|---|---|
| Calibrated. | Return the probe to the line. |
| Probe fouled. Clean it and retry. | Clean it, then Retry. |
| DI water has fluorescence. Use fresh DI water. | Zero only: the water is contaminated. Use fresh DI water. |
| Standard too high for the probe. / Standard too low for the probe. | Check the standard’s value and age, and that the value entered matches it. |
| No answer from the probe. Nothing changed. | Check the probe is connected, then Retry. |
The ST-710 calibrates pH. The ST-712 asks first: pH or ORP.
pH always runs three buffers, in this order: pH 7.00, then pH 4.00, then pH 10.00. You need the Pyxis pH 4-7-10 buffer pack or equivalent.

ORP (ST-712) uses one standard:
ST-712 ORP calibration is not yet tested on hardware. Check the reading against the standard after calibrating. The pH calibration above has been tested.
| Result | Do |
|---|---|
| Buffer value out of range. | The probe does not accept the buffer. Check it is the buffer the screen asks for. |
| Calibration failed. Check the buffer and probe. | Use fresh buffer, rinse, and Retry. A probe that keeps failing is at the end of its life — the ST-71x has no replaceable electrode. |
| No answer from the probe. Nothing changed. | Check the probe is connected, then Retry. |
A failed buffer is retried on its own. The buffers already passed stay done.
The CR-300 calibrates against a check cap, which stands in for a known corrosion rate. Two come with the probe: 2.0 mpy for steel, 0.1 mpy for copper.
CR-300 calibration is not yet tested on hardware. Until it is, check the probe against its cap without calibrating it: fit the cap, wait 15 minutes, and compare the reading with the cap’s value.

| Result | Do |
|---|---|
| Calibrated | Remove the cap and return the probe to service. |
| Current too low. Is the cap seated? | Reseat the cap, then Retry. |
| Current too high. Check the cap value. | Check the value entered matches the cap. |
| Standard out of range (0.02-100) | The value entered is outside what the probe accepts. |